3. Now select the “select” box on
service data as shown.
4. Now select the HEX file
required:-
QUATTRO Plus:
96101931QPLUSV2.0
QUATTRO Acute:
96101932QACUTEV2.0
5. Once this is achieved the HEX
file will appear in field indicated.
6. Now select the “Set pump ready
for download” box then read the
information screen and select
the “OK” box as shown.
7. Now select the “NO” box as
shown, and then within 5
seconds of selecting “NO” switch
the unit on.
Q U AT T R O
®
A C U T E & Q U AT T R O
®
P L U S
D H G - H E A LT H C A R E . C O M
2 5
3
7
5
4
7
6