S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
3 4
UPPER PUMP
MOUNT
T19R: PART
NUMBER
10371
3.4 Internal Pump and Upper Pump Post Replacement
1. Remove the rear case assembly via the 10 pozi-
drive screws (see section 3.2). The felt filter should
be changed during this procedure (section 3.1).
2. Withdraw the double pump unit, disconnecting
the air hose to the cycle control valve and the
power connector on the PCB . The pump is
retained only by the 4 anti-vibration mounts, so
it can be fully removed without the need for hand
tools.
3. Fit the new internal double pump with reverse
procedure.
4. If only the anti-vibration pump mount
is
damaged, replace any of the 4 off anti-vibration
mounts. The anti-vibration part simply pulls away
from mount support and new part can be push
fitted.
5. If one of the upper pump mounts is damaged
remove the damaged pump mount, unfasten the 2
x M4 x 12 pan head pozi screws and remove the
pump mount.
6. Fit the new upper pump mount using the existing
screws.
1
2
A
B
B
A
PUMP UNIT
INTERNAL
DOUBLE 230:50
T19R:
PART NUMBER
12163
PUMP UNIT
INTERNAL
DOUBLE 230:60
T19R:
PART NUMBER
12175
3
C
C
ANTI-
VIBRATION
MOUNT
KIT PACK
OF 4: PART
NUMBER
10984
4
D
5
6
D
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...