S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
4 0
5.
Detach the two tubes from the elbow or T-piece of the
air cell.
6. Remove the air cell from the centre popper strip.
Apply the reverse procedure for air cell replacement.
SEE SPARES LIST FOR APPROPRIATE AIR CELL
REQUIRED FOR MATTRESS.
7. If a T-piece connector is damaged or broken when
changing the air cell, it must be replaced. Extract old
connector from tubing or air cell and replace with new
connector.
8. If an elbow connector is damaged or broken when
changing the air cell, it must be replaced. Extract old
connector from tubing or air cell and replace with new
connector.
5
SEE SPARES LIST
FOR APPROPRIATE
ELBOW REQUIRED
FOR MATTRESS.
SEE SPARES LIST
FOR APPROPRIATE
ELBOW REQUIRED
FOR MATTRESS.
7
8
6
1.
Deflate the mattress, and un-fasten the air cells in the proximity of the CPR
(see section 3.10).
2.
Unfasten the 4 screws on the outside of the mattress . Ensure the Nyloc
nuts on the inside of the mattress are retained.
3. Slide the internal shroud off the CPR assembly to expose the tube mounts
and pull the tubes away from the four tube mounts on CPR back plate.
4. Remove the CPR assembly. Apply the reverse procedure for CPR
replacement.
3.11 Mattress CPR Replacement
CPR
ASSEMBLY
9MM G4 SPARE:
PART NUMBER
12189
CPR
ASSEMBLY
12MM SPARE:
PART NUMBER
10869
A
A
2
3
4
1
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...