1. Unfasten all the 10 x screws of the rear case,
comprising:
2 x M4 x 30 pan head pozi
8 x M4 x 50 pan head pozi
2. Remove the bottom case from unit. Then unfasten
the 4 self-tapping screws that are associated
with the handle or each of the handles that needs
replacing.
3. Replace the hanging handle assembly using the 4 existing screws and washers .
3.2 Handle Replacement
3.3 Rear Anti-vibration Pad T19R Replacement
1. Turn the QUATTRO unit so that the rear case
is facing up and clean the area where the old
anti-vibration pad was with a TECcare CONTROL
surface wipe (other wipes can be used) to clean
off any excess adhesive from the old pad.
2. Take the new anti-vibration pad and peel the
protector away from the adhesive side of the pad.
3. Position the new anti-vibration pad and apply to
the rear case apply force for 30 seconds to get a
good bond.
B
A
A
B
1
A
B
C
C
2
C
3
HANGING
HANDLE SET
SPARE: PART
NUMBER
10993
C
1
2
REAR ANTI-
VIBRATION
FEET SINGLE
T19R: PART
NUMBER
10181
3
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...