10. The unit will display ‘APPLY 160mm then press
DATA’. Set the sphyg. to 160mmHg and press the
information (data) key on unit.
11. The unit will now display the actual pressure that the
PCB is reading. Leave the unit in this mode and check
pressures.
S E R V I C E M A N U A L
D H G - H E A LT H C A R E . C O M
2 0
10
11
12
12. Set the sphyg. to 100mmHg and check that the unit displays 100mmHg. (+/-4mmHg).
13. Then set the sphyg. to 50mmHg to check that the unit displays 50mmHg (+/-4mmHg).
14. Set the sphyg. to 0mmHg to check that the unit displays 0mmHg (+/-4mmHg). The unit is now fully calibrated; the
power can be switched off and the mains cable disconnected.
13
14
2.6 Clearing Faults and Editing Information
1. Connect the unit to power but do not
switch on and connect the data lead
coming from USB communication box to
the unit and communication box to the
computer as shown in pictures below.
(LEAD COLOURS MIGHT BE DIFFERENT).
2. Switch the unit on with information
(data) key depressed simultaneously
to enter information menu.
3. Now using the service data USB v1.2
program you can transfer information to
and from unit. First select the “Device
Data” tab.
2
HOLD
1
3
Data
lead
Date lead
USB
cable
Содержание QUATTRO Acute
Страница 1: ...QUATTRO Acute QUATTRO Plus Service Manual DHG HE A LT HCAR E C OM ...
Страница 49: ...SERVICE MANUAL DH G HEALTHCARE COM 4 8 ...
Страница 50: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 4 9 ...
Страница 51: ...SERVICE MANUAL DH G HEALTHCARE COM 5 0 ...
Страница 53: ...SERVICE MANUAL DH G HEALTHCARE COM 5 2 4 5 Type 19R Circuit Diagrams ...
Страница 57: ...SERVICE MANUAL DH G HEALTHCARE COM 5 6 Notes ...
Страница 58: ...QU ATTRO ACUTE QUATTRO PLUS DH G HEALTHCARE COM 5 7 Notes ...
Страница 59: ...SERVICE MANUAL DH G HEALTHCARE COM 5 8 Notes ...