ASE 100 Operator’s Manual
5-22
Doc. 031842-01 5/02
9. Pull up the source needle assembly to remove the needle.
10. Insert the new source needle assembly (P/N 059923) into the top of the needle
block. Verify that the eye of the needle is toward the front of the needle block.
11. Reconnect the 1/4-inch stainless steel fitting to the needle block.
12. Reconnect the stainless steel tubing to the right side of the static valve.
13. Reinstall the right side panel:
a.
Hold the panel against the right side of the system. Engage the
small tabs on the top and bottom of the panel with the slots in the
chassis. Pull the panel forward until it locks into place.
b.
Replace the two Phillips screws on the inner front panel.
14. Carefully align the lower trim panel with the edges of the instrument, and then
gently push the panel into place.
15. Reinstall the waste bottle.
16. Turn on the ASE 100 main power switch.
17. Reconnect the gas source at the ASE 100 rear panel and turn on the gas
supply.
Be careful when handling the source needle. The tip of the
needle is sharp; in addition, it is fragile and can easily be bent.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...