2 • Description
Doc. 031842-01 5/02
2-9
2.1.6
Waste Bottle
A 250 mL collection bottle (P/N 056284, pkg. of 12) serves as the waste
bottle (see Figure 2-1) for the system.
Three vent lines, one from the pressure relief valve and two from the
needle mechanism, are connected to the waste bottle cap. The waste bottle
collects the small amounts of condensed solvent vented through these
lines.
A vent outlet line is connected to the waste bottle cap, also. Gas is vented
out this line to the rear panel of the ASE 100. For installation instructions,
see Section B.2.1.
NOTE Check the waste bottle regularly and empty
whenever necessary.
2.2
Rear Panel
Figure 2-4 illustrates the rear panel of the ASE 100.
_
The
POWER
switch provides on/off control of the main power for the system.
The power receptacle also includes the fuse holder. For instructions on how to
replace the fuses, see Section 5.11.
_
The oven voltage switches must be set to match the voltage from the power
source at the ASE 100 installation site. For instructions on how to set the
switches, see Section B.2.2.
_
The
NITROGEN
connector is connected to a nitrogen supply regulated to
between 1.03 and 1.38 MPa (150 to 200 psi). For installation instructions, see
Section B.2.1.
_
The
VENT
connector provides a connection for the vent outlet line. For
installation instructions, see Section B.2.1.
_
The model data label lists fuse and power information, as well as the ASE 100
serial number. You will be asked to provide the serial number when ordering
replacement parts for the system.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...