C • User Interface
Doc. 031842-01 5/02
C-13
C.2.5
Extraction Counters Screen
Use the
EXTRACTION COUNTERS
screen to check the number of
extractions the system has run or the number of pump strokes performed.
This can be helpful in scheduling the routine replacement of certain parts.
Figure C-12. Extraction Counters Screen
Parameter
Description
EXTRACTION
Indicates the number of extractions performed since
the ASE 100 was installed or the extraction counter
was last reset.
PUMP STROKE
Indicates the number of pump strokes performed
since the ASE 100 was installed or the pump stroke
counter was last reset.
EXTR. RESET
Select
RESET
to reset the extraction counter to zero.
PUMP RESET
Select
RESET
to reset the pump stroke counter to
zero.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...