5 • Service
Doc. 031842-01 5/02
5-33
11. Lift the lower AutoSeal tip out of the support tube located on top of the air
cylinder shaft (see Figure 5-17).
12. Note the indentation along the top edge of the support tube, where it meets the
lower AutoSeal tip (see Figure 5-17). Insert the tip of a small flathead
screwdriver into the indentation and twist the screwdriver just enough to raise
and loosen the lower tip.
13. Hold the solvent line securely with a 1/4-inch wrench, to prevent it from
moving; at the same time, use a 3/8-inch wrench to loosen the lower tip.
Remove the tip and discard it.
14. Screw the new AutoSeal tip (P/N 056641) onto the end of the solvent line.
While holding the solvent line securely with a 1/4-inch wrench, use a 3/8-inch
wrench to tighten the new AutoSeal tip.
15. Push the new AutoSeal tip into the support tube on top of the air cylinder.
16. Route the solvent outlet line through the vertical guide and reinstall the
Phillips screws in the guide.
Figure 5-17. Support Tube for Lower AutoSeal Tip
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...