5 • Service
Doc. 031842-01 5/02
5-19
7. The static valve is installed in the lower left corner of the component panel
(see Figure 5-8). Disconnect the orange tubing from the elbow press fitting on
the valve.
8. Disconnect the stainless steel line from the right side of the valve. (This line
connects to the source needle.)
9. Disconnect the stainless steel line (P/N 059821) from the top of the valve.
(This line connects to the lower AutoSeal tip.)
10. Loosen the Allen hex screw (9/64-inch) on the clamp ring below the valve
base.
11. Lift the old valve off the valve base.
12. Lower the new static valve (P/N 048778) onto the valve base. Tighten the
Allen hex screw on the clamp ring.
13. Reconnect all of the lines.
Figure 5-8. Connections to the Static Valve
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...