5 • Service
Doc. 031842-01 5/02
5-5
5.3
Removing the Right Side Panel
Many parts that the user can service (the pump check valves, etc.) are located
behind the panel on the right side of the ASE 100.
To remove the right side panel:
1. Turn off the ASE 100 main power switch.
2. Unscrew the waste bottle from the bottle cap and set the bottle aside.
3. Toggle the needle switch to the
UP
position. If a collection bottle is installed,
remove the bottle from the bottle holder and set it aside.
4. Turn off the gas supply. Disconnect the gas source at the ASE 100 rear panel.
5. Remove the lower trim panel of the ASE 100:
a.
Stand directly in front of the ASE 100, with both hands facing
palm upwards.
b.
Hook the first two fingers of each hand under the top edges of the
bottle holders, near the
WASTE
and
COLLECTION
labels (see
c.
Brace your thumbs against the bottom of the cell door and the
control panel. Pull forward until the top of the trim panel pops off
the ball studs that secure it to the system.
Figure 5-3. Removing the Lower Trim Panel
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...