5 • Service
Doc. 031842-01 5/02
5-17
9. Disconnect the stainless steel line from the left side of the valve. (This tubing
connects to the upper AutoSeal tip.)
10. Disconnect the green PEEK waste line from the top of the valve.
11. Loosen the Allen hex screw (9/64-inch) on the clamp ring below the valve
base.
12. Lift the old valve off the valve base.
13. Lower the new relief valve (P/N 048889) onto the valve base. Tighten the
Allen hex screw on the clamp ring.
14. Reconnect all of the lines.
15. Reinstall the right side panel:
a.
Hold the panel against the right side of the system. Engage the
small tabs on the top and bottom of the panel with the slots in the
chassis. Pull the panel forward until it locks into place.
b.
Replace the two Phillips screws on the inner front panel.
16. Carefully align the lower trim panel with the edges of the instrument, and then
gently push the panel into place.
17. Reinstall the waste bottle.
18. Turn on the ASE 100 main power switch.
19. Reconnect the gas source at the ASE 100 rear panel and turn on the gas
supply.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...