ASE 100 Operator’s Manual
3-12
Doc. 031842-01 5/02
3.1.7
Rinsing/Priming the System
During a rinse cycle, approximately 5 mL of solvent is pumped through
the system. Run a rinse cycle at the following times:
_
After initial setup
_
After refilling the solvent bottle
_
After changing solvents (rinse twice to remove all of the previous
solvent)
_
When the solvent lines contain bubbles
_
After a run is aborted
_
Before shutting down the system, if an extraction used acidic solvents
or basic solvents (see Section 3.8)
_
After a shutdown of more than one day (see Section 3.8)
Before starting a rinse cycle:
_
Check that the rinse cell size is matched to the extraction cell size.
_
Adjust the cell holder to accommodate the rinse cell, if necessary.
_
Install the rinse cell in the cell holder.
_
Close the cell door.
When you are ready, press the
RINSE
button to start the rinse cycle. For
detailed instructions, refer to Section B.2.6.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...