Meter view
The power supply's front panel features a high contrast 4.3-inch LCD color display. Press [Meter View].
Item
Description
1
Remote sense status
2wire or 4wire indicates that 2-wire or 4-wire sensing measurement is in use.
2
Preferred mode status
Volt or Curr indicates the preferred mode for output on or output off transitions.
3
Current range indicator
Displays the current range indicator. This indicator will only show when low current range is selected.
4
Output status
OFF: The output is off
CV: The output is in constant voltage mode
CC: The output is in constant current mode
CCN: The output is limited by negative constant current.
UR: The output is unregulated
OV: Over-voltage protection is tripped
OC: Over-current protection is tripped
OT: Over-temperature protection is tripped
+IPK: The output is protected by positive peak current
-IPK: The output is protected by negative peak current
5
Instrument/Interface status
: The instrument is connected to USB.
: The instrument is connected to LAN.
(blinking): The instrument is under identification mode via
remote interface.
: The instrument is not connected to LAN.
(blinking): The LAN connection is at fault.
ARB: Arb is running.
!ERR: An error has occurred (press Utilities > Error to view the
Error Log).
6
Ratings and protection
Displays the present over-voltage protection (OVP) setting.
7
Output delay
Displays the OCP, Output On, and Output Off delay status/values.
8
Voltage slew
Displays the rising slew and falling voltage slew rate.
9
Soft front panel menu
Keysight E36150 Series User's Guide