Pulse Arb (Option E36150ADVU)
Step 1– Select Arb waveform
Press Arb to access the Arb selection menu. Then select Pulse.
Press Back to return to Arb main menu.
Step 2 – Configure the Pulse Arbs
Press Properties. Then program the following parameters:
Configure the Pulse Arb accordingly. Refer to the below table for details.
Parameter
Available Key Settings
Description
Arb Type
Voltage or Current
Sets the output Arb type.
Default is Voltage.
Action required:
Press Arb Type to toggle between Volt and Curr.
Volt
Selects voltage as output Arb type.
Curr
Selects current as output Arb type.
Start Setting (I0 or
V0)
minimum to maximum
Sets the setting before and after the pulse.
Pulse Setting (I1
or V1)
minimum to maximum
Sets the amplitude of the pulse.
Delay (t0)
0 – 3600 s
Sets the delay time after the trigger is received but before the pulse starts.
Pulse Width (t1)
0 – 3600 s
Sets the width of the pulse.
End time (t2)
0 – 3600 s
Sets the time the output remains at the end setting after the pulse completes.
Keysight E36150 Series User's Guide
93