Trigger Settings >
Trigger
Source
V1 Level, I1 Level,
Run Key, Arb Key,
On/Off, DIO or
Remote
Configures the trigger source.
Default is Run Key.
Action required:
Press Trigger Source to select a trigger source. This trigger
source will trigger the data logging on the output channel.
V1 Level
Selects the voltage trigger level as
trigger source.
Triggers the measurement when the
voltage of the corresponding output
passes through the specified level.
I1 Level
Selects the current trigger level as
trigger source.
Triggers the measurement when the
current of the corresponding output
passes through the specified level.
Run Key (Run/Stop Key) Selects the Datalog Run Stop key as
a trigger source.
Arb Key (Arb Run/Stop
Key)
Selects the Arb Run Stop key or [List
Run/Stop] as a trigger source.
On/Off (Output On/Off
Key)
Selects the Output On/Off key as a
trigger source.
DIO (DIO Trigger In)
Selects any configured digital IO
(with Trigger Input function) as a
trigger source.
Remote (Remote
Command)
Selects a remote interface
command as a trigger source.
Level
Specifies a trigger level if you selected a Voltage level or Current
level as the trigger source. Along with the level, you must also
specify a slope.
Slope
Specifies if the measurement will be triggered on the positive
(up-slope) or negative (downslope) portion of the waveform.
Trigger Positions Specifies a trigger offset. This allows the specified percent of pre-
trigger data to be logged to the file. The trigger position is
expressed as a percentage of the data log duration. For
example, if you specified a data log duration of 30 minutes and a
trigger position of 50%, the Data Logger will log 15 minutes of
pre-trigger data to the file before the trigger occurs. Sub-
sequently, 15 minutes of post-trigger data will then be logged to
the data file.