Lg MinMax
Off or On
Sets to enable or disable the Log Min/Max checkbox. Enabling
the checkbox will log the minimum and maximum values to the
data log file. With Log Min/Max checked, the resulting file size
will triple.
Default is Off.
Action required:
Press Lg MinMax to toggle between Off and On.
From the remote interface:
To enable current or voltage data logging:
SENS:DLOG:FUNC:CURR 1
SENS:DLOG:FUNC:VOLT 1
You cannot data log output power from the remote interface. To obtain power data, you need to data log both
voltage and current and then calculate the power from the resulting voltage and current data.
To log the minimum and maximum values to the data log file for the enabled output:
SENS:DLOG:FUNC:MINM 1
To specify a data log of 1000 seconds on the enabled output:
SENS:DLOG:TIME 1000
To specify a sample period of 400 milliseconds between data samples on theenabled output:
SENS:DLOG:PER 0.4
To send an immediate trigger signal to the data logger:
TRIG:DLOG
To select the immediate trigger source (triggers the Data Logger immediately when initiated):
TRIG:DLOG:SOUR IMM
To select the rear panel trigger input (all connector pins that have been configured as trigger sources):
TRIG:DLOG:SOUR EXT
To select a BUS trigger source:
TRIG:DLOG:SOUR BUS