Settings
Available Key Settings
Description
Trigger Source
Key, IO, or Rmt
Sets the trigger source for the system.
Default is Rmt.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (Arb Run/Stop Key)
Selects the Arb Run Stop key or [List Run/Stop] as
a trigger source.
IO (DIO Trigger In)
Selects any configured digital IO (with Trigger Input
function) as a trigger source
Rmt (Remote Command)
Selects a remote interface command as a trigger
source.
Trigger Delay
0 – 3600 s
Sets the trigger delay in seconds.
Default is 0 s.
Repeat Count
1 to 9999 (standard)
1 - 16,777,216 (Option
E36150ADVU)
Sets the Arb repeat count. This sets the number of times that an Arb is executed before it
completes.
Default is 1.
Continuous
-
Enables check box to repeat the Arb continuously.
Step 4 – Run the output sequence list
When Arb List is initiated, all the List properties cannot be configured, which includes adding or removing step
from the list.
When Step mode transient system is initiated, properties such as voltage mode, current mode, trigger source,
and trigger delay cannot be configured.
If both Voltage and Current Mode is set to Fix and Trigger Source is set to Key, pressing
Run
will automatically
set both voltage and current mode to Arb and start the list operation.
– Set Voltage Mode to Arb: Press V Mode Arb.
– Set Current Mode to Arb: Press I Mode Arb.
– Set Trigger Source to Key. Press Trig Src Key.
– Press the color-coded [On] key to enable the output.
– Press Run to start the Arb operation. To abort the operation, press Stop. Alternatively, press [List Run/Stop].
– Press Back to exit and return to the previous menu.
– Press [Scope/Datalog] to enter Scope view and display the Arb.
80
Keysight E36150 Series User's Guide