Utilities Menu
Utilities provides the following features:
Store and recall state and power on settings
View USB and LAN status as well as configure LAN, GPIB (optional), and digital IO
Displays the instrument's error queue
Manage files, including creating screen shots
Utilities Menu - Store and Recall State
The Utilities menu is shown below.
Store/Recall saves and recall states. In general, state files store volatile settings associated with measurements.
The instrument states include volatile settings such as:
– Voltage, current, OVP, OCP delay, and OCP delay start
– Voltage slew, output preference and output sense
– Output state
– Operation mode (Independent)
– Output on/off sequencing
– Arb settings
– Trigger settings
– Digital I/O output data and bus setting
– Data logger settings
– Scope settings
Keysight E36150 Series User's Guide
143