Commissioning
6.4 Commissioning via PDM
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
47
Note
Measured Track Values
The "Measured Track Values" should show stable values within the normal range, i.e. the
gains should be stable values between 3 and 12 (smaller sizes low values and larger sizes
high values) and the actual velocities for the paths should be stable, constant and smoothly
changing values between 0 and 10 m/s.
Store settings on the PC
Store the device settings after verifying the values.
●
Choose: "Read"
→ "Upload to PC" to get all the settings.
● Store the complete settings on your PC via "File" → "Export".