Electrical connection
5.3 Compact and remote transmitters: Connecting output and power supply
32
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
1. Connect power supply to L1, N and protective earth (PE).
Figure 5-2
Output and power supply connection
2. Fix the power supply cable to the terminal housing with the clamp.
3. Ground transmitter and sensor.
Note
Cable straps
It is recommended to use cable straps.
Note
Sensor grounding
To ensure identical potential for sensor and transmitter, a direct grounding of transmitter and
sensor is recommended.
Step 2: Wiring pulse output
CAUTION
Pulse output
The pulse output must be connected to equipment complying with Low Voltage Directive in
order to be considered safe. The isolation within the transmitter pulse output is only a
functional isolation.
1. Prepare the cable ends as shown in the figure above.
2. Connect output cables to terminals 56 and 57 (Output A) and to 66 and 67 (Output B)
according to use.
The output functions are used as shown in the figure above.
3. Ground the cable shielding by fixing it with the clamp.
Содержание SONO 3500 CT
Страница 1: ...Ultrasonic Flowmeter Operating Instructions SONO 3500 CT 07 2019 Edition ...
Страница 2: ......
Страница 8: ...Table of contents 6 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 14: ...Introduction 1 5 Further Information 12 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 38: ...Electrical connection 5 5 Sealing of SONO 3500 CT 36 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 50: ...Commissioning 6 4 Commissioning via PDM 48 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 54: ...Functions 7 4 Hardware key 52 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 64: ...Service and maintenance 8 7 Battery disposal 62 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...