Electrical connection
5.3 Compact and remote transmitters: Connecting output and power supply
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
31
5. Connect grounding wire plug to the grounding terminal (PE) on the PCB.
Note
Sensor grounding
To ensure identical potential for sensor and transmitter, a direct grounding of transmitter
and sensor is recommended.
6. Tighten both transducer cable glands to ensure they are firmly sealed (recommended
torque: approx. 20 Nm).
7. Make sure that the transducer cables are not overbent. Minimum cable bend radius is 50
mm.
5.3
Compact and remote transmitters: Connecting output and power
supply
The wiring of output and power supply is done in three steps:
1. Wiring power supply
2. Wiring pulse output (if relevant)
3. Mounting transmitter
Step 1: Wiring power supply
WARNING
Power supply requirements
Make sure the power supply requirements stated on the nameplates are met!
Содержание SONO 3500 CT
Страница 1: ...Ultrasonic Flowmeter Operating Instructions SONO 3500 CT 07 2019 Edition ...
Страница 2: ......
Страница 8: ...Table of contents 6 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 14: ...Introduction 1 5 Further Information 12 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 38: ...Electrical connection 5 5 Sealing of SONO 3500 CT 36 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 50: ...Commissioning 6 4 Commissioning via PDM 48 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 54: ...Functions 7 4 Hardware key 52 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 64: ...Service and maintenance 8 7 Battery disposal 62 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...