Installing/Mounting
4.2 Sensor installation
20
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
Make sure that flowmeter is positioned as low as possible to prevent air from being trapped
in flowmeter at transducers.
Find a position on the pipeline where inlet pipe to flowmeter has a straight length as
specified below.
Single bend
1 x 90° bend
L2: Min. 10 x pipe diameter
L1: 3 x pipe diameter