CyberOptics Corporation
63
Main Electrical Panel
Replacing the Fuses
Parts
Fuse 10 A
Fuse 6 A
➢
To replace the fuse
1. On the rear of the machine, turn the AC main circuit breaker to OFF.
2. Verify that AC main power is off and locked off. Use a voltmeter to verify that the power is
off.
3. Pull out the faulty fuse and replace with a new fuse of the same rating.
4. Turn the AC main circuit breaker to ON.
i
Note:
For Parts Replacement, contact CyberOptics for part numbers.
Warning:
Hazardous voltages may be present. Make sure that power is disconnected from the
mains before replacing the fuse.
Caution:
Recommended that only trained service personnels perform this operation.
Table 11.
Fuse Listings
Circuit Breaker
Rating
Purpose
CB2
6.0 A
To protect the processor outlet
CB3
6.0 A
To protect PS1
CB4
6.0 A
To protect PS2
CB5
6.0 A
To protect PS3
CB6
10.0 A
To protect X-Axis servo amplifier
CB7
10.0 A
To protect Y-Axis servo amplifier
CB8
6.0 A
To protect monitor outlet