97 | R-net Technical Manual | Chapter 3 – Programming
9.11
ACTUATOR ENDSTOP TIMEOUT
This sets the amount of time the actuator motor current has to be above the ends top detection level (or actuator current limit if
Endstop Autodetect is set to No) before the drive is cut. It is actually very rare to have to adjust this parameter, but it c an be
useful to prevent nuisance stopping, i.e. false detections of endstop.
The programmable range is 0 milliseconds to 500 milliseconds in steps of 10 milliseconds.
10
BATTERY
10.1
LOW BATTERY ALA RM
This sets whether there will be an audible alarm to signal a low battery condition. The p oint at which the alarm will sound
corresponds to the Low Battery Flash Level setting.
The programmable options are Yes and No.
10.2
LOW BATTERY FLA SH LEVEL
This sets the point at which the battery gauge on the LCD screen starts to flash slowly to warn of a low battery condition.
The programmable range is 1 to 10 in steps of 1, and corresponds to the number of bars shown on the battery gauge. For
example, if this value is set to 2, then the flashing will occur when the gauge drops to 2 bars.
10.3
CABLE RESISTANCE
This calibrates the system to compensate for the electrical resistance of the battery cable and connectors to the Power Modul e.
This is necessary to ensure an accurate estimation of battery charge.
The programmable range is 0mOhms to 500mOhms in steps of 1mOhm.
The value should take into account the electrical resistance in both the positive and negative paths, including connectors.
10.4
CALIBRATION FACTOR
This allows further fine calibration of the battery gauge. This is normally set at the factory and should not need further
adjustment.
10.5
LOW VOLTAGE CUT-OUT, LOW VOLTAGE TI ME
These two parameters all ow a low battery voltage cut-out function to be programmed. Such a function can be used to prevent
over-discharge of the batteries.
If the battery voltage is below the value set by Low Voltage Cut-out for a period of time longer than that set by Low Voltage
Time, then drive will be prevented. This will be recorded in the Power Module’s error log.
Drive will then be permitted to resume when the battery voltage rises above the value of Low Voltage Cut-out.
The programmable range for Low Voltage Cut-out is 16V to 22V in steps of 0.5V.
Содержание R-NET
Страница 1: ...PG DRIVES TECHNOLOGY R NET TECHNICAL MANUAL SK77981 14 ...
Страница 46: ...Curtiss Wright PG Drives Technology SK77981 14 46 R net Technical Manual Chapter 2 Installation 6 1 3 CRIMPING ...
Страница 58: ...Curtiss Wright PG Drives Technology SK77981 14 58 R net Technical Manual Chapter 2 Installation ...
Страница 174: ...Curtiss Wright PG Drives Technology SK77981 14 174 R net Technical Manual Chapter 8 ISM 8 ...
Страница 194: ...SK77981 14 194 R net Technical Manual Chapter 10 Diagnostics Curtiss Wright PG Drives Technology ...
Страница 206: ...SK77981 14 206 R net Technical Manual Chapter 12 Specifications Curtiss Wright PG Drives Technology ...