31
Instruction Manual
7.3 Access Levels and User Modes
Corning® Ascent™ FBR system has three access levels: Default, Operator, and Engineer. When entering the username and password, a
keyboard will pop up (Figure 7-3). Table 7-3 shows the access at each user level.
Figure 7-3.
Login pop-up box and keyboard.
Default Login:
When the system restarts or when the user logs out by selecting the padlock icon, the system reverts to Default login
which only allows view access to the process without change access.
Operator Login:
The Operator login allows access to basic functionality, but not to manually controlling components (such as pumps and
pinch valves), VPN, or other advanced features.
Engineer Login:
At Engineer level, the user can manually override control and equipment modules setpoints.
Table 7-3.
Login access levels.
Access Levels
Action
Default Operator Engineer
Observe process overview screens
Sys button for software information and clean screen
Observe historical trends
Sys button for user interface shutdown
Initiate and Cancel Phases
Global Abort button use
Change probe calibration
Pump calibration
Pressure zero point adjustment during Initialize Phase
Data button (Data File Entry/Meta-data Entry/Recirc Volume)
Change Date and Time
View Alarm History
Password Admin button to reset operator and engineer passwords
USB file access via Data button
Set Operator Mode for manual control of devices (pumps, pinch valves, gas controllers) and control loops
Edit device and control loop properties (Names/Units/Alarms/PID Loop Constants)
Diagnostics button
Change alarm setpoints
Table 7-4 shows the operation modes to perform either manual or semi-automated sequences for a flexible experiment.
Table 7-4.
Operation modes.
Mode
Description
Program
The system determines the loop/device outputs.
Operator
The user determines the loop/device outputs.
Auto
The control loop input is the setpoint in engineering units.
Manual
The control loop input is set at a fixed percentage output.