3.8 Maintenance and cleaning of the allPIXA pro camera
During operation of the device, particles such as dust etc. may be settled on the optical
components (lens) of the camera. These deposits affect the optical image and the function of
the camera negatively.
Chromasens recommends regular inspection and cleaning. The cleaning
intervals depend on the actual operating and ambient conditions (for
example dust-laden atmosphere).
3.8.1 Cleaning intervals
Cleaning intervals depend on the environment. Regular inspection and cleaning intervals must
be specified depending on the degree of soiling.
3.8.2 Cleaning process
Body of the allPIXA pro camera heats up during operation.
Before cleaning, you have to switch off the device. Always allow hot
surfaces to cool down before cleaning the device.
The device works with electric power. Before cleaning the device, make
sure that the device is disconnected from the power supply.
All surfaces requiring cleaning can be wiped with a soft, lint-free cloth which can be moistened
with Isopropanol.
Never use any other liquid or cleaning agent than those stated in this manual.
Never use hard or sharp tools for cleaning the device.
Inspect the device to ensure that cleaning was effective and repeat, if necessary.
If it is not possible to clean a component due to irremovable contamination, it must be
replaced.
3.9 Disposal
The device consists of different kinds of material and for its disposal the materials have to be
separated according to the local regulations. The material has to be disposed properly to avoid