Chromasens CD40127_R02 allPIXA_pro_User Manual.docx
104
5
Start the shading correction:
In the
Data set
list, select the data set, in which the offset reference should be
generated and stored:
In the
Timeout value
box, enter the timeout value in seconds.
Move the target slightly or defocus the lens to avoid streaky images, if necessary.
Note:
Any features on the target surface (for example dust, scratches) end up in the
calibration profile of the camera. To avoid this if you use a static (non-moving) white
reference target, use a clean white ceramic or plastic material, not paper. Ideally, the white
object should move during the calibration process because the movement results in an
averaging process and the camera diminishes the effects on any small variation in the white
reference.
Click
Next
.
6
Wait for result
A progress bar is shown while the reference is created and stored. When it has finished
successfully, the wizard shows the following message:
To close the wizard and to restore the camera settings, click
Next
.