9.4.3 Enable continuous white control
Name
Name
Enable continuous white control
Function
Enables the closed-loop control for white balancing
Channels
all
Unit
Bool
Dependency
None
Notes
Camera:
Value:
Function:
Lower limit
all
0
Off
Upper limit
all
1
On
9.4.4 Position and mode for white reference mark
Name
Name
Position and mode for white reference mark
Function
It is possible to set the reference zone either to front tap, rear tap, to both
taps independently or to automatic detection.
It is strongly recommended to use automatic detection. In this case, the
camera chooses automatically which tap is used as master.
Changing this default behavior should only be done by experienced users.
Channels
All
Unit
Integer
Dependency
Only takes effect if the
Enable continuous white control
check box is
selected, or during a white reference operation if you click
Perform white
reference
on the toolbar.
Notes
For more information about the exact size and position, see section 9.3
This function is only available, if
User level
is set to
Guru
.
Camera:
Function:
Lower limit
All
Front TAP data are used for gain control. Rear
channels follow (slave rear)
All
Rear TAP data are used for gain control. Front
(slave front)
All
Front and Rear TAP data are controlled
independently
Upper limit
All
Automatic master tap detection out of area position