Q9 Digital Color Doppler Ultrasound System
84
r
4i
:Radius of i-th ellipse(4CH)
LVLd: Length of the left-ventricular long axis at end diastole for 2CH or 4CH, whichever is longer.
LVLs: Length of the left-ventricular long axis at end systole for 2CH or 4CH, whichever is longer.
(3) EDV=
π
Errore. L'origine riferimento non è stata trovata.2i
2
(4) ESV=
π
Errore. L'origine riferimento non è stata trovata.2i
2
ri: Radius of the i-th circle(2CH)
LVLd: Left ventricular long-axis length at end diastole
LVLs: Left ventricular long-axis length at end diastole
(5) EDV=
π
Errore. L'origine riferimento non è stata trovata.4i
2
(6) ESV=
π
Errore. L'origine riferimento non è stata trovata.4i
2
ri: Radius of the i-th circle(4CH)
LVLd: Left ventricular long-axis length at end diastole
LVLs: Left ventricular long-axis length at end diastole
<<Items to be measured>>
Meas. item name
Description [Unit]
Meas. method
EDV
(
A2C
)
End-diastolic volume (A2C )[ml]
Measurement (trace)
ESV
(
A2C
)
End-systolic volume (A2C) [ml]
Measurement (trace)
EDV
(
A4C
)
End-diastolic volume (A4C) [ml]
Measurement (trace)
ESV
(
A4C
)
End-systolic volume (A4C) [ml]
Measurement (trace)
<< Items to be calculated >>
Calc. item name
Description [Unit]
Calc. formula
SV
(
SimpBP
)
Stroke Volume[mL]
SV=EDV – ESV
EF
(
SimpBP
)
ejection fraction
EF=SV/EDV
3.
Bullet volume
This method calculate the left ventricular volume using the 2D-mode long-axis image and the short-axis image at the
level of the mitral valve.
Calculation formula for volume
EDV=Errore. L'origine riferimento non è stata trovata.
ESV=Errore. L'origine riferimento non è stata trovata.
<<Items to be measured>>
Meas. item
name
Description [Unit]
Meas. method
LVAMd
Left ventricular short-axis area at the level of the mitral
valve at end diastole.
Refer to “Area-trace
measurement”
LVLd
Left ventricular long-axis length at end diastole.
Refer to "Distance
Measurement
(Distance)".
LVAMs
Left ventricular short-axis area at the level of the mitral
valve at end systole.
Refer to “Area-trace
measurement”
LVLs
Left ventricular long-axis length at end systole.
Refer to "Distance
Measurement
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...