Q9 Digital Color Doppler Ultrasound System
56
Blood Effection
Choose different blood effection
Smooth, HighRes
Shortcut Adjustment
:
Shortcut
(
action
)
Menu
Menu Function
Menu Item adjustable Rang
SK1
(
press
)
Frequency
Adjust the frequency of
probe
Refer to current probe
SK1
(
rotate
)
Steer
Adjust angle of sampling box
of blood flow under the linear
probe
Adjust range from -20 to 20
SK2
(
press
)
Wall Filter
Change of Wall filtering
0~3 adjustable
SK2
(
rotate
)
Color Map
Change of Color Map
Adjust option:User, Type:1- 9
SK3
(
press
)
Flow
Change sensitivity of blood
High, Middle and Low for option
SK3
(
rotate
)
PRF
Adjust PRF value
The maximum PRF depends on the probe and the
position of sampling box.
SK4
(
press
)
Invert
Realize invert of blood flow
Turn on or Turn of the invert
SK4
(
rotate
)
Persistence
Improve the current color
0~7 grade adjustable. Adjust range: 1.
SK5
(
press
)
Density
Change images density
High and Low adjustable.
SK5
(
rotate
)
Baseline
Change the baseline of color-
map
-3~3 adjustable.
Other adjustments
:
1
、
CFM gain control.
Rotate C- knob after C lights is on .Adjust the color gain and adjust range is from 0 to 255.
2
、
Sampling box.
Moves sampling box through trackball to the area which you are interested and press the ENTER- key to confirm the
position of sampling box then adjust the size of box trough trackball and press ENTER- key to confirm.
Use the STEER toggle key to change the direction of sampling box under the linear probe.
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...