Q9 Digital Color Doppler Ultrasound System
77
same steps to create other distance pairs
6) Press the DIST-key or EXIT-key to exit.
Distance ratio
In B mode, press the DIST-key to bring up the first white plus sign (“+”) cursor.
Rotate SK1,choose %stD
。
Refer to the distance measurement method, measure the distance of second line.
After finishing the measurement, the distance of the two lines will appear automatically on the screen and
calculate the ratio of the two lines automatically.
Area—trace
1) In B mode, press the TRACE-key to bring up the trace start cursor plus sign (“+”).
2) Use the trackball to move the cursor, press ENTER-key to set the start point.
3) Use the Trackball to trace the perimeter of the object to be measured, and press the Enter-key to set the end
point. The end point and the starting point of the trace will be automatically connected with a straight line .The
area and the Circumference of the shape will be calculated. With the movement of the cursor, the result will
appear on the right side of the screen, and will change along with the moving of the trace.
4) To get the multiple data, repeat the same steps to create other distance pairs.
5) Press the Trace-key or EXIT-key to exit.
Area—ellipse
1) In B mode, press the TRACE-key to bring up the trace start cursor plus sign (“+”).
2) Rotate SK2
,
choose ellipse.
3) Move the Trackball and press ENTER-key to set start point
4) The second point marker is displayed by a line connecting the first and second points. Use the Trackball to
lengthen the line in order to change the diameter and press the ENTER-key to fix the length.
5) Use the Trackball to change the length of another axis.Press the ENTER-key to fix the length. The total area
and circumference will be displayed in the result window.
6) To get the multiple data, repeat the same steps to create other distance pairs.
7) Press the Trace-key or EXIT-key to exit.
Area ratio
1) In B mode, press the TRACE-key to bring up the trace start cursor plus sign (“+”).
2) Rotate SK2
,
choose %stA.
3) Refer to the area-ellipse method to measure two ellipses.
4) After finishing the measurement, the area of the two lines will appear on the screen, and the area ratio will be
calculated automatically.
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...