Q9 Digital Color Doppler Ultrasound System
155
Only water-solvable solutions or gel can be used. Petroleum or mineral oil-based materials may harm the
cover.
When the transvaginal probe is activated outside patient’s body, its acoustic output level should be decreased
to avoid any harmful interference with other equipment.
Operation Procedure:
Put on medical sterile glove
Get the condom for the package.
Unfold the condom.
Load some ultrasound gel into condom.
Take the condom with one hand, and put the probe head into the condom.
Fasten the condom on the end of the probe handle.
Confirm the integrity of the condom, and repeat the above steps to the condom if any damage to the condom is
found.
8.3.3 Cleaning and Disinfecting TV Probe
We strongly recommend wearing gloves when cleaning and disinfecting any endocavitary probe.
Every time before and after each exam, please clean the probe handle and disinfect the transvaginal probe
using liquid chemical germicides
If the probe is contaminated with body fluids, you should disinfect the probe after cleaning.
Regard any exam waste as potentially infectious and dispose of it accordingly.
CAUTION:
Since the probe is not waterproof, you should disconnect it from the system before cleaning or disinfecting.
Before and after each exam, please clean the probe handle and disinfect the transvaginal probe using liquid
chemical germicides.
Cleaning
You can clean the transvaginal probe to remove all coupling gel by wiping with a soft cloth and rinsing with
flowing water. Then wash the probe with mild soap in lukewarm water. Scrub the probe as needed and use a soft
cloth to remove all visible residues from the transvaginal probe surface. Rinse the probe with enough clean
potable water to remove all visible soap residues, and let the probe air dry.
CAUTION:
Please remove the cover (if any) before cleaning the probe.(The cover like condom is one time
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...