Q9 Digital Color Doppler Ultrasound System
69
Service: Click the drop list to select storage function.
AE title: Name the server at will
Host: Set the IP of the service host
Port: Make the port of the work station the same as the port of the server.
Alias: The name of current link
Time out: time delay
Mode: batch, follow, manual selectable
Batch: all the saved images in current exam will be sent to the server in one time
Follow: When one image is saved, the system will send this image to the server.
Manual: The user has to select the image by manual.
After all the settings are completed, please click OK to save current setting. Click clear to clear all the setting.
Clicking cancel will not save the current setting.
DICOM storage
Batch mode: In batch mode, when the patient finishes his exam, press End key to end the current exam. All
the saved images will be saved as DCM format and sent to the server automatically.
Follow: In follow mode, during the exam, press
key to save the image. The current image will be
saved as DCM format and sent to the server.
Manual: In manual mode, the user has to select the image by manual and do DICOM transfer. Press
key to enter into review interface. Select the images need to be sent, press
key to send.
4.
DICOM print
DICOM print parameters setting.
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...