Q9 Digital Color Doppler Ultrasound System
13
8
0.5754
0.1903
0.0631
0.0158
I’=I*R
F
Where I’ is the intensity in soft tissue, I is the time-averaged intensity measured in water.
Tissue Model:
Tissue temperature elevation depends on power, tissue type, beam width, and scanning mode. Six models are
developed to mimic possible clinical situations.
Thermal Models Composition Mode
Specification
Application
1
TIS
Soft tissue
Unscanned
Large aperture (>1cm
2
)
Liver PW
2
TIS
Soft tissue
Unscanned
Small aperture (<1cm
2
)
Pencil Probe
3
TIS
Soft tissue
Scanned
Evaluated at surface
Breast color
4
TIB
Soft tissue and bone
Scanned
Soft tissue at surface
Muscle color
5
TIB
Soft tissue and bone
Unscanned
Bone at focus
Fetus head PW
6
TIC
Soft tissue and bone
Unscanned/scanned
Bone at surface
Transcranial
Soft tissue:
Describes low fat content tissue that does not contain calcifications or large gas-filled spaces.
Scanned: (auto-scan)
Refers to the steering of successive burst through the field of view, e.g. B and color mode.
Unscanned:
Emission of ultrasonic pulses occurs along a single line of sight and is unchanged until the transducer is moved to a
new position. For instance, the PW, and M mode.
TI:
TI is defined as the ratio of the In Situ acoustic power (W.3) to the acoustic power required to raise tissue temperature
by 1°C (Wdeg), TI = W.3/Wdeg.
Three TIs corresponding to soft tissue (TIS) for abdominal; bone (TIB) for fetal and neonatal cephalic; and cranial
bone (TIC) for pediatric and adult cephalic, have been developed for applications in different exams.
An estimate of the acoustic power in milli-watts necessary to produce a 1°C temperature elevation in soft tissue is:
Wdeg = 210/fc,
for model 1 to 4, where fc is the center frequency in MHz.
Wdeg = 40 K D
for model 5 and 6, where K (beam shape factor) is 1.0, D is the aperture diameter in
cm at the depth of interest.
MI:
Cavitation is more likely to occur at high pressures and low frequencies in pulse ultrasound wave in the tissue, which
contains the bubble or air pocket (for instance, the lung, intestine, or scan with gas contrast agents). The threshold
under optimum conditions of pulsed ultrasound is predicted by the ration of the peak pressure to the square root of the
frequency.
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...