Q9 Digital Color Doppler Ultrasound System
4
Do not remove the protective covers on the system. These covers protect users against hazardous voltages.
Cabinet panels must remain in place while the system is in use. A qualified electronic technician must make all
internal replacements.
Do not operate this system in the presence of flammable gases or anesthetics.
All peripheral devices (unless certified as medical grade) that are connected to the system must be powered
through the electrical outlet with an optional isolation transformer.
Suggest power off the system in 30 minutes if the system continuously works in 8 hours.
Notice upon Installation of Product
Separation distance and effect from fixed radio communications equipment: field strengths from fixed transmitters,
such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast transmitter cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the ultrasound system is used exceeds the applicable RF compliance level as stated in
the immunity declaration, the ultrasound system should be observed to verify normal operation. If abnormal operation
is observed, additional measures may be necessary, such as re-orienting or relocating the ultrasound system or using an
RF shielded examination room may be necessary.
Use either power supply cords provided by or designated by CHISON. Products equipped with a power source
plug should be plugged into the fixed power socket which has the protective grounding conductor. Never use any
adaptor or converter to connect with a power source plug (e.g. three-prong-to-two-prong converter).
Locate the equipment as far away as possible from other electronic equipment.
Be sure to only use the cables provided by or designated by CHISON. Connect these cables following the
installation procedures (e.g. wire power cord separately from signal cables).
Lay out the main equipment and other peripherals following the installation procedures described in this manual
.
Notice against User Modification
The user should never modify this product.
User modifications may cause degradation in Electrical Safety. Modification of the product includes changes in:
Cables (length, material, wiring, etc.)
System configuration/components
User modifications may cause degradation in EMC performance. Modification of the product includes changes in:
Cables (length, material, wiring, etc.)
System installation/layout
System configuration/components
Securing system parts (cover open/close, cover screwing)
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...