Keep a log of all probe maintenance, along with a picture of any probe malfunction.
WARNING:
The probes are designed to be used only with this ultrasound system. Use of these probes on any
other system or a non-qualified probe may cause electric shock or damage on the
system/transducer.
8.2.2 Cleaning and Disinfecting
CAUTION:
These transducers are not designed to withstand heat sterilization methods. Exposure to
temperatures in excess of 60 º C will cause permanent damage. The transducers are not designed
to be totally submerged in fluid, as permanent damage will result if the entire transducer is
submerged.
Probe Safety
Handling precautions
Ultrasound probes are highly sensitive medical instruments that can easily be damaged by improper handling. Use
care when handling and protect from damage when not in use. DO NOT use a damaged or defective probe.
Failure to follow these precautions can result in serious injury and equipment damage.
Electric shock hazard:
The probe is driven with electrical energy that can injure the patient or user if live internal parts are contacted by
conductive solution:
DO NOT immerse the probe into any liquid beyond the level indicated by the immersion level diagram. Never
immerse the probe connector into any liquid.
Prior to each use, visually inspect the probe lens and case area for cracks, cuts, tears, and other signs of
physical damage. DO NOT use a probe that appears to be damaged until you verify functional and safe
performance. You need to perform a more thorough inspection, including the cable, strain relief, and connector,
each time you clean the probe.
Before inserting the connector into the probe port, inspect the probe connector pins. If a pin is bent, DO NOT
use the probe until it has been inspected and repaired/replaced by a CHISON's authorized service engineer.
Electrical leakage checks should be performed on a routine basis by CHISON's authorized service engineer.
Mechanical hazard:
Содержание Q9
Страница 126: ...Q9 Digital Color Doppler Ultrasound System 120 Input Template Input annotation from templates ...
Страница 143: ...Q9 Digital Color Doppler Ultrasound System 137 Click Yes Select Dynamic Mode click Install ...
Страница 145: ...Q9 Digital Color Doppler Ultrasound System 139 Click NEXT Select Local printer attached to this computer and Click Next ...
Страница 173: ...Q9 Digital Color Doppler Ultrasound System 167 ...