91
Solution:
Check the cable outlets for loose contacts at both ends.
2. Problem:
The screen remains black after starting the system (does not start).
Cause:
Battery discharged or damaged, or the fuse is blown.
Solution:
Recharge the battery and re-start the system. If the problem persists, refer to Section 8
dealing with fuse replacement or contact a Cheetah Medical service person.
3. Problem:
Burned out pixels on the LCD screen
Cause:
LCD malfunction
Solution:
Please contact a Cheetah medical representative for service.
4. Problem:
LCD failure during monitoring
Cause:
DC power supply to the inverter or inverter failure
Solution:
Please contact a Cheetah medical representative for service.
5. Problem:
Buzzer failure upon power on
Cause:
Buzzer turned off or buzzer malfunction
Solution: