66
Special Note
During the first 130 seconds of operation the system
performs an automatic calibration function and a corresponding message is
posted on the screen. At this time, for optimal results, the patient should lie or
sit as quietly as possible, preferably without speaking or making sudden
movements. Once the initial calibration is completed the message is taken off
the screen, signaling that standard monitoring can resume. In the event of
excessive patient movement or talking during this period, the system will filter
away interferences however this may prolong the initial calibration period
beyond 130 seconds.
7. During the measurement, the screen display can be changed and the measurement
can be paused as explained in section 2.5.
8. In order to take an NIBP measurement, connect the air hose by pushing it in and
turning it clockwise. Make sure that the NIBP is set to 'Enable' on the NIBP Options
settings screen. When the Starling SV is set, place the cuff around the patient's arm,
over bare skin above the elbow. Lift the patient's arm to the level of the heart. In case
that the Starling SV is set to Automatic NIBP operation, the Starling SV
will inflate the cuff at a given time according to the settings that were selected for
NIBP prior to beginning the test (see chapter 4) and will display the pressure on the
screen. A manual measurement can be taken at any given time.
Note
The arm is to be passively raised to the level of the heart; this should not
be done by the patient. Moreover, the patient should be as relaxed as possible
throughout the NIBP measurement.
9. At the end of the measurement, in order to completely end the test, select the Stop
option from the screen's pop-up menu or by clicking the OK button on the
STARLING SV. Then select the Back button on the Starling SV in order to return to
the Menu screen.