![Cheetah Medical Starling SV Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/cheetah-medical/starling-sv/starling-sv_user-manual_2595119029.webp)
28
Note
It is possible to completely mitigate or significantly alleviate the problem by
placing the Starling SV sensor 2.5 inches (approximately 6 cm) or more away from
the external pacemaker percutaneous lead. Unipolar internal pacemakers are rarely
used nowadays.
1.15
SpO2
Introduction:
Pulse oximetry is a non-invasive method allowing the monitoring of the oxygen
content of a patient's hemoglobin.
A sensor is placed on a thin part of the patient's body, usually a fingertip or
earlobe, or in the case of an infant, across a foot. Light with red wavelengths and
light with infrared wavelengths is sequentially passed from one side to a photo
detector on the other side. Changing absorbance of each of the two wavelengths
is measured, allowing determination of the absorbances due to the pulsing
arterial blood alone, excluding venous blood, skin, bone, muscle, fat, and (in
most cases) fingernail polish. Based upon the ratio of changing absorbance of the
red and infrared light caused by the difference in color between oxygen-bound
(bright red) and oxygen-unbound (dark red or blue, in severe cases) blood
hemoglobin, a measure of oxygenation (SpO2 - the percent of hemoglobin
molecules bound with oxygen molecules) can be made.
Populations that may use the SpO2 function:
The Starling SV SpO2 module is intended for use with neonatal, pediatric, and
adult patients during both no motion and motion conditions and for patients
who are either well or poorly perfused, dependent on the SpO2 sensor used.
SpO2 Sensors for the NELL-1 SpO2 Module:
The NELLCOR DS100 SPO2 sensor comes supplied with the STARLING SV
with the SpO2 functionality. In addition, the Starling SV with NIBP and SpO2
functionality is also compatible with any of the following NELLCOR SpO2
sensors that have the following accuracy specifications (*Arms):
Sensor
Adults/Pediatrics
Accuracy
Neonates
Accuracy
REUSABLE:
D-YS (Infant to Adult)
± 3 digits
± 4 digits
D-YS with D-YSE Ear
Clip
± 3.5 digits
N/A
D-YS with D-YSPD
SpotClip
± 3.5 digits
N/A
DS-100A
± 3 digits
N/A
OXI-A/N
± 3 digits
± 4 digits
OXI-P/I
± 3 digits
N/A
DISPOSABLE:
OxiCliq A:
± 2.5 digits
N/A
OxiCliq P:
± 2.5 digits
N/A
OxiCliq N (adult):
± 2.5 digits
± 3.5digits