![Cheetah Medical Starling SV Скачать руководство пользователя страница 88](http://html.mh-extra.com/html/cheetah-medical/starling-sv/starling-sv_user-manual_2595119088.webp)
87
or
These messages appear with an additional message:
Reason – disconnection between the Starling SV device and the patient's body. This
results in a suboptimal Starling SV signal.
Cause:
Patient's cable or one or more of the Starling SV Sensors are not connected
properly to the Starling SV device or to the patient.
Solution:
Check that the Starling SV right or left (according to the message) sensors are well
attached to the patient body and connected properly to the Starling SV cable.
Check the connection of the Starling SV cable to its socket in the rear panel on
the back side of the Starling SV device. Sometimes earlier sensor replacement is
warranted, especially if the sensors are wet or if the underlying skin is hairy. If the
problem persists, stop the test and start a new one.
3. Message:
Reason – This message will only appear at the end of a Study period for 5 seconds
when the device is unable to report reliable results at the end of a Study due to long
periods of excessive movements or long periods of sensor disconnections.
Cause:
Patient's cable or one or more of the Starling SV Sensors were not connected
properly to the Starling SV device or to the patient or/and the device detected
patient or cable movements that distorted excessively the signal.
Solution:
The user should discourage excessive cable or patient movements in Rest type
tests or minimize as much as possible cable movements when performing a test
in Stress test mode. The user should be alerted to sensor or cable disconnections
and repair the problem if occurs as soon as possible.
4. Message: