is set to Enabled. When the NIBP is set to disabled, the NIBP is not functional and
will not respond to NIBP measurements requests.
4.11
NIBP Operation Mode
This option allows the user to choose between manual NIBP measurements, which
means the operator can take a measurement at any given time according to his
patient's needs, and AutManual measurements, which means that the
operator can perform a measurement at any given time and the STARLING SV will
perform an automatic NIBP measurement at a specific period of time (every 30mins
for example), according to the Time Interval settings that were set by the operator.
While using the NIBP Aut Manual option, if a manual measurement is taken
2 minutes prior to the beginning of an automatic one, the automatic measurement will
be delayed. (Should the time interval be set to 30 minutes, the next automatic
measurement will take place 32 minutes after the manual measurement has been
taken).
Note
While actively measuring blood pressure, the NIBP activity timer will appear
on the NIBP box to state that the NIBP module is being used.
Special Note
The NIBP, MAP, CP, CPI, TPR, TPRI and DO2I are not as
frequently refreshed as the rest of the STARLING SV's parameters. Once a
parameter has not been refreshed for a period of ten minutes, its color will begin to
fade out by a gradual color fade-out process that takes another ten minutes to
complete. The parameter will regain its full display color only when a new
measurement is taken.
4.12
NIBP Time Interval
This option allows the user to set the NIBP time interval for the Aut Manual
option, so that an automatic NIBP measurement will take place at a pre-set period of
time according to his patient's needs.
Note
The time interval icon appears on the NIBP box to state the time left before
the next automatic measurement.
4.13
NIBP Initial Inflation Pressure
The initial inflation pressure system default is set to 160 mmHg.
This option allows the user to set the NIBP initial inflation pressure to a different
threshold level from the system's default, according to the patient's systolic blood
pressure. The options to choose from are (in mmHg): 120, 140, 160, 180 and 200.
Setting the correct initial inflation pressure reduces the incidence of both over-