NIBP module is not recognized. NIBP functionality will be disabled. Please contact a
Cheetah Medical support representative.
Cause:
The NIBP module is disconnected from the device or it stopped functioning.
Solution:
Please contact a Cheetah medical representative
5. Message:
Phase Calibration Error. Unable to calibrate. Please check cable and sensor
connections and try again
Cause:
The sensors are not applied appropriately or are detached from the patient
body. The sensors may need to be replaced. Alternatively, the patient cable is
disconnected either from the device or from the sensors.
Solution:
Replace sensors if you suspect suboptimal sensor application. Check that the sensors
are well placed and connected to the patient cable. Check that the patient cable is
securely connected to the back of the monitor.