CAUTION
Neonates measurements must always use a 3 meters patient hose
in order to avoide overpressure errors caused by a lack of air volume within the
overall pneumatic system.
CAUTION
A compressed or kinked connection hose may cause continuous cuff
pressure resulting in blood flow interference and potentially harmful injury to the
patient.
CAUTION
Too frequent BP measurements can cause injury to the patient due to
blood flow interference.
CAUTION
Pressurization of the BP cuff can temporarily cause loss of function of
simultaneously used monitoring equipment on the same limb
WARNING
Regarding the effect of blood flow interference and resulting harmful
injury to the PATIENT caused by continuous CUFF pressure due to connection
tubing kinking.
WARNING
Regarding the need to check (for example, by observation of the limb
concerned) that operation of the AUTOMATED SPHYGMOMANOMETER does
not result in prolonged impairment of the circulation of the blood of the patient.
Adverse Reactions
Allergic exanthema (symptomatic eruption) in the area of the cuff may result,
including the formation of urticaria (allergic reaction including raised edematous
patches of skin or mucous membranes and intense itching) caused by the fabric
material of the cuff.
Petechia (a minute reddish or purplish spot containing blood that appears on the skin
surface) formation or Rumple-Leede phenomenon (multiple petechia) on the forearm
following the application of the cuff, which may lead to Idiopathic thrombocytopenia
(spontaneous persistent decrease in the number of platelets associated with
hemorrhagic conditions) or phlebitis (inflammation of a vein) may be observed.
Cuff Selection & Placement
It is important to select the cuff size that is appropriate to the diameter of the patient's
upper arm. Use the Range Lines on the inside of the cuff to determine the correct size
cuff to use.
Wrap the cuff around the arm making sure that the Artery Marker is aligned over the
brachial artery (as shown in Figure 1.3). If possible, do not wrap the cuff over the
patient's clothing. The cuff should fit snug to the patient's arm for maximum
oscillometric signal quality. An appropriate sized cuff should be placed on the non-