22
1.7
Bedside Care
To ensure that the Starling SV signals will be recorded to an adequate
standard, it is important to follow the below guidelines:
Starling SV Sensor Placement
1) Shave body hair and wipe the patient’s skin with alcohol (70%) if necessary.
If alcohol is used to prep the skin, it is advised to let the skin dry prior to
placement of the STARLING SV sensors in order to preserve optimal gel
performance. Finally, use the skin-prep “sand paper” provided in the sensor
set to roughen the skin.
Place the STARLING SV Sensors on the patient’s skin as seen bellow (Figure 1.4)
Figure 1.4 Starling SV Prewired Sensor Placement
Note When the sensors are misplaced or become detached the screen will display a
warning that will notify the user that there is loss of signal or poor signal quality.
CAUTION
Do not use Starling SV sensors after the expiration date.
CAUTION
In the unlikely event that the patient has an allergic reaction to the
STARLING SV sensors, detach the sensors immediately from the patient and avoid
any further placement of Starling SV sensors on this patient.
1.8
Label
The following label is attached to the Cheetah STARLING SV device: