The module meets all relevant parts of the following Safety
Standards:
IEC60601-1:1997 2nd Edition
IEC60601-1:2005 3rd Edition
IEC/EN60601-2-30:1999/2000
IEC 80601-2-30:2009
AAMI SP10:2002(R)2008
ISO 81060-2:2009
EN1060-1:1996+A2:2009
EN1060-3:1997+A2:
Error Code List & Definitions
CAUTION
Wen an NIBP error occurs, measurement values will still be reported.
These measurement values should be disregarded when an error code is reported.
Exceeded retry count
Corrective Action:
• The patient may have been moving too much.
• Check that the cuff is properly tightened.
• Check that the cuff is in the correct position.
• Check that the correct size cuff is being applied.
• Check that there is no excessive clothing between the arm and the cuff.
Exceeded measurement time limit
Corrective Action:
• The patient may have been moving too much.
• Check that the cuff is properly tightened.
• Check that the cuff is in the correct position.
• Check that the correct size cuff is being applied.
• Check that there is no excessive clothing between the arm and the cuff.
Artifact / Erratic Oscillometric Signal
Corrective Action:
• The patient may have been moving too much.
• Check that the cuff is in the correct positions.
• Check that the correct size cuff is being applied.