39
corrected to the patient’s BSA. Once selected, a small 'V' sign will appear on the left
hand side of the 'Index Display' option on the pop-up menu
Time Scale -
this option enables the user to create wider or narrower timescale view,
for example in order to focus on a short period of interest or to visualize many hours
of monitoring and observe a longer-term trend. The user has a choice of different
time-scale options. This can also be done by moving two fingers apart/towards each
other on the touch screen when trends are displayed.
Session Controls
Initiate Calibration -
This option enables the user to run a new patient calibration
process on the system while a test is ongoing.
Start –
by choosing this option the user will begin running a new test. Another way to
do so is by clicking the knob while on the main Run screen as long as a Test is not
already running at the time.
Pause
– this option allows the user to halt a running test for an unlimited period of
time, without terminating it. The test can be resumed at any time by choosing the
Continue option on the Test Controls' sub menu. When pause is operated, “pause”
indicator will appear in the alerts bar.
Stop -
this option prompts a pop-up screen to confirm the end of the test by the user.
Silence Alarm -
this option allows the user to silence the alarm for two minutes
during a test.
Protocols
The Protocol Wizards guide the user in step-by-step fashion through Starling SV
supported protocols. The Protocol Wizard is designed to simplify administration and
data management related to hemodynamic tests such as the Passive Leg Raise, or
Bolus injection. Once chosen, the tests perform automatically without interruption of
monitoring and will only request user involvement and confirmation where required.
The wizard will track the relevant clinical parameters and will also calculate additional
clinical information to provide the user with the end point of the protocol. A protocol
can be initiated at any stage during a test, except for the calibration phase. Simply
press the Menu button on the Starling SV to access the screen's pop-up menu or just
select the Menu with the touch screen. In addition, the user can start the PLR and the
Bolus Protocols directly from the Challenge stage. If this is selected, the system will
automatically analyze the last 3 minutes of readings to establish whether or not there
is a valid Baseline stage. The system will not allow for a direct challenge stage if the
last 3 minutes readings contain poor signals or excessive movements. When selecting
the 'Protocols' option a sub menu opens with a choice of the following protocols:
PLR test, and Bolus. Once selected, the protocol wizard will appear on screen and