play icon will now be replaced by the pause icon). In order to browse between the
different types of display screens, select the Switch Display option from the Run
screen's menu, then use the knob to select the display screen of interest or just swipe
you finger vertically on over the screen until the required screen is displayed. In order
to access the Run pop-up menu press the Menu button, then use the knob to navigate
between the options.
To view the quality of the data generated during the test the raw data and the
STARLING SV and STARLING SV average morphology are displayed in real-time.
In addition, the user can customize the parameter boxes to be displayed by simply
touching the parameter and selecting the desired parameter from the menu.
The menu is accessed from the various different Run screens (Run, Dashboard,
Trends and Numerical Display) enable the following options:
NIBP
NIBP Time Interval -
this option allows the user to change the time interval settings
during an ongoing test. Once selected, a small "V" sign will appear to the left of this
option. The change is relevant for that specific test only. The NIBP time interval for
the following test will be set according to the pre-defined default as appears on the
Settings screen.
This option is available through all Run screens' pop-up
menus.
Manual NIBP -
this option allows the user to start a manual NIBP measurement.
The NIBP will initiate the measurement immediately after selection. Following the
manual reading the system will continue to periodically obtain an NIBP measurement
according to the default time interval chosen in the Settings screen.
End NIBP -
this option allows the user to quit all NIBP measurements immediately,
which also deflates the cuff. The change is relevant for that specific test only. The
NIBP measurements for the following test will be set according to the pre-defined
default as appears on the settings screen.
This option is available through all Run
screens' pop-up menus.
Resume NIBP –
this option replaces the End NIBP option on the sub menu. It
only becomes available after the End NIBP option was selected and is meant to re-
enable NIBP measurements during a running test.
View
Y Axis Auto Scaling -
this option activates an automatic focus on the dynamic area
of interest vs. the whole spectrum of potential results. Thus, this option provides
more insight on trends.
Index Display –
by selecting this option, the CO, SV, CP and TPR parameters will
be replaced by their indexed parameters on the display. Indexed parameters are
Содержание Starling SV
Страница 1: ...R IFU 05 Revision 3 Starling SV User Guide ...
Страница 10: ...9 Figure 1 2 Back panel of the Starling SV Figure 1 3 Front and side of the Starling SV ...
Страница 46: ...45 Figure 2 10 Starling SV Tabular display screen ...
Страница 86: ...85 6 23 DBP SBP stands for Diastolic Blood Pressure ...