13
For neonatal patients, blood pressure measurements made with the Advantage OEM
BP Module Series are equivalent to those obtained by intra-arterial blood pressure
devices within the limits prescribed by ANSI/AAMI SP10: 1992 & 2002 (mean error
difference of ± 5 mmHg or less, standard deviation of 8 mmHg or less).
The automatic selection between the 3 NIBP population modes is performed when
the entering the patient demographic information as follows:
Adult
– Age above 12 years (Automatic, default)
Pediatric
– Weight above 7Kg and age 12 and below (Automatic)
Neonate
– Weight 2.5Kg and below (Automatic)
NOTE
The user shall be given the option to select Neonate or Pediatric NIBP
population mode if the patient weight is between 2.6Kg and 7Kgs.
NOTE
in Neonate mode the following ranges apply:
Systolic
40-130mmHg
MAP
26-110mmHg
Diastolic
20-100mmHg
Initial Inflation Pressure
90mmHg (default)
Variable from 60 to 140mmHg
Warnings & Precautions
This device should not be used when oscillometric pulses may be altered by other
devices or techniques such as External Counterpulsation (ECP) or Intra Aortic
Balloon Pump Counterpulsation.
DO NOT
use the NIBP module for any purpose other than specified in this
manual without written consent and approval from Cheetah Medical.
DO NOT
use the device in the presence of flammable gaseous anesthesia agents
because of flame hazard.
DO NOT
attach the cuff to a limb being used for IV infusions or any other
intravascular access, therapy or an arterio-venous (A-V) shunt. The cuff inflation can
temporarily block blood flow, potentially causing harm to the patient.
DO NOT
apply the BP cuff over a wound as this can cause further injury.
DO NOT
For patients that have had a mastectomy, cuff should be applied to
opposite arm.
CAUTION
Substitution of a component different from that supplied may result in
measurement error. Repairs should be undertaken only by personnel trained or
authorized by Cheetah Medical. Do not modify this equipment without authorization
from Cheetah-Medical.