ERROR CODES
37
FIRMWARE
The firmware contains error codes that can detect the root location of a malfunction. These data are less useful for the
troubleshooting service. However, if one of these errors is encountered, the error code targets the unit that can be replaced
or serviced (remote / module). To see the last error code, switch off the remote control. Then press and maintain the two
channel keys in the middle (keys #2 and #3) and press the On/Off button.
Code
Description / Comment
Part to change
0xEA63
Mi-scan error: seems to occur more on channel 3 & 4. Corrected bug since the 0105 module version.
Module FW
0x00D2
Seems particular out of range. handling of topstim from Remote control when out of range + start/stop
-
0x5002
Stimulation was in pause state instead of stop
Remote
0x4902
Remote control cannot connect with the module. Missing feedback from the Rfnode master in the given
timeout value
Remote FW
0x4905
Remote freed command state machine, but command state machine not ready
Remote
Control
0x00C8
Module hardware failure not detected during self tests, probably due to hardware failure
Module HW
0x4782
A pause was sent, but a stimulation module did not send its status to the remote in the correct timing.
This might be an out of range side-effect
-
0x0115
ERROR_SC2_IStimNotCompensated
Module (80/20
FW vs HW)
0x00CE
This might be an out of range side-effect
-
0x0113
Pulse duration too long, more than 4.5ms between trigger-pulse and end-of-pulse. Check what's
happening when stop is received exactly during pulse
Module FW
0x5006
Mostly probably missing one of the 32bits frame when load program to flash memory
0x00D0
V_CMD not relative to PWM output. Check what's happening when stop or emergency stop is received
exactly during pulse
Module (80/20
FW vs HW)
0x0116
ERROR_SC8_IStimOverAbsoluteMax. Over current
-
0x480B
RF chip is not working at 100%. RF communication is ok, but no topstim is generated
Remote
(FW/HW)
0x47E0
Check if this error could have been raised before buy. If not, hide the Menu RD. (1 month elapsed
between error and first program done. Did the error occur during prod, and associated to the matching
user?)
Remote FW
0x025A
Not enough time given to dc power to generate HV before trigger pulse was called. Remove this error,
simply stop HV generation
Module FW
0x2001
Wrong menu ID. The function asserting this error is called with flash-based value. It's near not possible
that this value to be the one stored in record
Remote FW
Figure 10 : Error code screen
Note: Final accessories may differ from the picture